Description
- Automatic internal adjustment in the case of a change in temperature ≥ 0,5 °C or timecontrolled every 4 h, guarantees high degree of accuracy and makes the balance independent of its location of use
- Dosage aid: High stability mode and other filter settings can be selected
- Simple recipe weighing and documenting with a combined tare/print function. In addition, the ingredients for the recipe are numbered automatically and printed out with their corresponding number and nominal weight
- Identification number: 4 digits, printed on calibration protocol freely programmable
- Printout of a GLP-compliant calibration report conveniently at the touch of a button
- Automatic data output to the PC/printer each time the balance is steady
- Large glass draught shield with 3 sliding doors for easy access to the items being weighed.
- Protective working cover included with delivery
Semi Micro balance
ABT 100-5NM
Product family: ABT
The premium model with single-cell weighing system
TECHNICAL DATA
Weighing capacity [Max]: 101 g
Readability [d]: 0,00001 g
Verification scale interval [e]: 0,001 g
Unit: sg, ct
Conformity assessment possible ex works
Minimum weight [Min]: 0,001 g
Linearity: ± 0,00015 g
Reproducibility: 0,00005 g
Stabilization time: 10 s
USP Minimum weight of sample (k = 2, U = 0.1%): 100 mg
Interfaces: RS-232 standard
Adjustment options: Adjusting with external weight, Internal calibration (automatic), Adjusting with internal weight (manual)
Ambient temperature1: 0 °C – 30 °C
Dimensions weighing surface : (Ø)80 mm
Material draught shield: glass
Material housing: metal
Dimension housing (W×D×H): 217×356×338 mm
Material weighing plate: stainless steel
Net weight : approx.8 kg
Construction type of the scale: Single-range balance