Features
● Integrated with Probe N: the eddy current principle to measure the thickness of insulating coatings on non-magnetic materials.
● Zero point calibration and two-point calibration to correct the system error of the probe
● Features two working modes: DIRECT and BATCH & two measuring ways: CONTINUE and SINGLE
● Statistics include the mean, maximum, minimum, test numbers and standard deviation.
● Memory of 500 data
● Deletion of current data, calibrated data, limit data and all data stored.
● Integrated with printer to print the statists values if needed
● Low battery indication and error alarm
● Buzz during operation for indication
● Backlight for the screen
● Auto or manual shutdown
Probe types : N
Working principle ; Eddy current
Measuring range : 0 to 1250 μm
Minimum resolution : 0.1μm
Tolerance :
Zero point calibration : ±(3%H + 1.5)
Two-point calibration : ±[(1%-3%)H + 1.5]
Measuring condition :
Min. curvature radius (mm) : Convexity 3; concave 10
Min. testing area diameter (mm) : Ø5
Critical thickness of substrate (mm) : 0.3
Power supply : Battery AAA 1.5V (2pcs)
Working temperature : 0~40°C
Dimensions (mm) : 145×60×28
Weight (g) : 132