Features
.High-speed ARM+DSP processor with low power consumption, strong computing power and fast calculation speed, and the result will be available within 1 second after the travel.No need to wait.
.Collect thousands of datas in the distance of travel to accurately reproduce the surface of the workpiece.
.Precise digital filter. Ensure machine repeatability is less than 2%.
.10% accuracy error in line with the national standard.
.Adapt to high temperature, high humidity, high vibration working environment.
.The test results can be traced back to the roughness parameters of the national benchmark.
.Diamond touch needle, strong and durable.
.Unique software processing. Start the test immediately, no need to wait.
.Precision inductive bridge sensor with resolution up to nanometer level.
.Low measuring force sensor in line with national standards has low impact on user workpiece.
Characteristics
.320*240 Highlight color LCD screen, exquisite display of roughness waveform.
.320*240 Highlight color LCD screen.Adapt to the low-light workshop environment
.Simple nine-button operation. clear and simple.
.Simple and intuitive menu design is convenient to set roughness parameters.
.Lithium battery power supply with USB charging. Charge it once, users can use it thousands of times.
.Very low motor sound. Strong operating comfort.
.Simple software calibration. Errors can be quickly corrected on site.
.Up to 26 roughness parameters.
.With the USB communication function, users can upload the data to the computer and PLC.
.100 sets of test data storage.
TECHNICAL SPECIFICATIONS
Application : Measure surface roughness
Type : Portable
Measuring range :
Z-axis(vertical) : Upper limit:200μm; Lower limit:0.010μm;
X-axis(horizontal) : 17.5mm(0.71inch)
Circuit resolutionZ-axis(vertical) : 0.002μm±20μm; 0.004μm±40μm; 0.008μm±80μm;
Parameter : Ra Rz Rq Rt;Rp Rv R3z R3y Rz(JIS);Rs Rsk Rku Rsm Rmr;
Standards : ISO4287; ANSI B46.1;DIN4768; JIS B601;
Display : 320*240 LCD color display. Display TP curve, roughness filter waveform;
Communication : USB interface
Storage : 100 sets of data
Filter : RC,PC-RC,Gauss,D-P;
Sampling Length (mm)(lr) : 0.25,0.8,2.5;
Evaluation Length (mm)(ln) : Ln=lr×n n=1~5;
Sensor :
Measurement principle : Electromagnetic inductance;
Stylus : Natural diamond, 90° tip radius, radius of stylus point 5μm;
Test force : <4mN;
Skid : Cemented carbide, drive direction radius 40mm;
Drive speed : lr=0.25 ,Vt=0.135mm/S; | lr=0.8 , Vt=0.5mm/S; | lr=2.5 , Vt=1mm/S;|On returning : Vt=1mm/S;
Display resolution ratio : 0.001μm;
Display error : ±(5nm+0.1A)
Repeatability of displayed value : ≤3%;
Residual profile : ≤0.01μm;
Battery : 3.7V rechargeable Li-on battery